Skip to content

A randomized clinical trial of therapeutic hypothermia mode during transport for neonatal encephalopathy.

CPQCC Publication
TitleA randomized clinical trial of therapeutic hypothermia mode during transport for neonatal encephalopathy.
Publication TypeJournal Article
Year of Publication2015
AuthorsAkula VPriya, Joe P, Thusu K, Davis AS, Tamaresis JS, Kim S, Shimotake TK, Butler S, Honold J, Kuzniewicz M, DeSandre G, Bennett M, Gould J, Wallenstein MB, Van Meurs K
JournalJ Pediatr
Volume166
Issue4
Pagination856-61.e1-2
Date Published2015 Apr
ISSN1097-6833
KeywordsAsphyxia Neonatorum, Body Temperature, Brain Diseases, Female, Follow-Up Studies, Humans, Hypothermia, Induced, Infant, Newborn, Infant, Newborn, Diseases, Intensive Care Units, Neonatal, Male, Prognosis, Transportation of Patients
Abstract

OBJECTIVE: To determine if temperature regulation is improved during neonatal transport using a servo-regulated cooling device when compared with standard practice.

STUDY DESIGN: We performed a multicenter, randomized, nonmasked clinical trial in newborns with neonatal encephalopathy cooled during transport to 9 neonatal intensive care units in California. Newborns who met institutional criteria for therapeutic hypothermia were randomly assigned to receive cooling according to usual center practices vs device servo-regulated cooling. The primary outcome was the percentage of temperatures in target range (33°-34°C) during transport. Secondary outcomes included percentage of newborns reaching target temperature any time during transport, time to target temperature, and percentage of newborns in target range 1 hour after cooling initiation.

RESULTS: One hundred newborns were enrolled: 49 to control arm and 51 to device arm. Baseline demographics did not differ with the exception of cord pH. For each subject, the percentage of temperatures in the target range was calculated. Infants cooled using the device had a higher percentage of temperatures in target range compared with control infants (median 73% [IQR 17-88] vs 0% [IQR 0-52], P < .001). More subjects reached target temperature during transport using the servo-regulated device (80% vs 49%, P <.001), and in a shorter time period (44 ± 31 minutes vs 63 ± 37 minutes, P = .04). Device-cooled infants reached target temperature by 1 hour with greater frequency than control infants (71% vs 20%, P < .001).

CONCLUSIONS: Cooling using a servo-regulated device provides more predictable temperature management during neonatal transport than does usual care for outborn newborns with neonatal encephalopathy.

DOI10.1016/j.jpeds.2014.12.061
Alternate JournalJ. Pediatr.
PubMed ID25684087
Grant ListUL1 RR025744 / RR / NCRR NIH HHS / United States
UL1 TR000093 / TR / NCATS NIH HHS / United States